HDT Vicat Tester
HDT Vicat Tester Plastic Testing Equipment Material Testing Kuala Lumpur (KL), Malaysia, Selangor, Penang Supplier, Suppliers, Supply, Supplies | Redmark Industry Sdn Bhd

Description:

This testing machine is primarily used to determine the heat distortion temperature and Vicat softening point temperature of nonmetal materials such as plastic, rubber, nylon, hard insulating materials, long fiber reinforced materials, and high-strength thermosetting laminated materials. It features easy operation, precise control, and high intelligence.

Design Standards:

  • GB/T 1633
  • GB/T 8802
  • GB/T 1634
  • ISO 2507
  • ISO 75
  • ISO 306
  • ASTM D1525
  • ASTM D648

Specifications:

  • Temperature Range: Room Temperature to 300¡æ
  • Maximum Temperature Indication Error: 0.1¡æ
  • Temperature Indication Precision: ±0.5¡æ
  • Maximum Deformation Indication Error: ±0.001 mm
  • Deformation Measurement Range: 0 to 10 mm
  • Sample Holder Number: 3 sets
  • Weight of Load Bar and Tray: 68 g
  • Test Span Range: According to Standard (64 or 100 mm)
  • Heating Medium: Methyl silicone oil or transformer oil (conventional flash point is 260¡æ, high temperature type provided by the customer)
  • Cooling Method:
    • Natural cooling for temperatures above 150¡æ
    • Water or natural cooling for temperatures below 150¡æ
  • Heating Power: 3 kW
  • Supply Voltage: 220V, 50Hz

Characteristics:

  • PC Control: Uses PC computer control for advanced testing.
  • Digital Deformation Measurement: Utilizes Far Easton dial indicators for precise digital deformation measurement.
  • Real-Time Data Processing: Real-time curve drawing and secondary data treatment directly on the PC.
  • Data Export: Test data can be imported into OFFICE software for editing and typesetting.

Specification 

 

Temperature range Room Temperature—300¡æ
Maximum temperature indication error 0.1¡æ
Temperature indication precision ±0.5¡æ
Maximum deformation indication error ±0.001mm
Deformation measurement range 0—10mm
Sample holder number 3 sets
Weight of load bar and tray 68g
Test span range According to Standard (64 or 100mm)
Heating medium

Methyl silicone oil or transformer oil

(The conventional flash point is 260 ¡æ, the high temperature type needs to be provided by the customer).

Cooling method:

More than 150 ¡æ(natural cooling)

less than 150¡æ (water or natural cooling)

Heating power 3kw
Supply voltage: 220V, 50Hz

This testing machine offers a comprehensive solution for determining the heat distortion and Vicat softening points of various nonmetal materials, ensuring high precision and ease of use for laboratory and industrial applications.